Laporkan Masalah

Rancang Bangun dan Validasi Alat Uji Sensitivitas Antibiotik Semi-Otomatis pada Vibrio parahaemolyticus

Arman Faries, Indah Istiqomah, S.Pi, M.Si, Ph.D.; Ir. Prapto Nugroho, S.T., M.Eng., D.Eng., IPM.

2026 | Tesis | S2 ILMU PERIKANAN

Pengujian sensitivitas antibiotik metode difusi cakram yang dilakukan secara manual memiliki keterbatasan dalam hal standardisasi teknis, akurasi dan efisiensi operasional. Penelitian ini bertujuan untuk merancang, mengevaluasi kinerja mekanis, dan memvalidasi alat uji sensitivitas antibiotik semi-otomatis berbasis metode difusi cakram standar EUCAST untuk mendeteksi resistensi Vibrio parahaemolyticus terhadap Tetracycline, Oxytetracycline, dan Enrofloxacin. Sistem mengintegrasikan modul otomatis untuk penyiapan media, standardisasi suspensi bakteri, inokulasi, dan penempatan cakram antibiotik. Kinerja alat dievaluasi dengan membandingkannya terhadap metode manual berdasarkan akurasi volume cairan, pembacaan densitas bakteri, homogenitas kultur, sebaran bakteri pada media agar, serta ketepatan penempatan cakram antibiotik. Validasi dilakukan menggunakan 19 strain Vibrio parahaemolyticus terhadap Oxytetracycline, Tetracycline, dan Enrofloxacin (171 titik uji), dengan parameter ukuran zona hambat dan interpretasi sensitivitas antibiotik, mengacu pada sistem standar (manual). Hasil pengujian menunjukkan pompa peristaltik memiliki kesalahan volume < 1 xss=removed>Tetracycline (p-value = 0,509) dan Enrofloxacin (p-value = 0,803), serta perbedaan rata-rata minimal sebesar -0,85 mm pada Oxytetracycline (p-value = 0,015). Hasil numerik tersebut selaras dengan validasi fungsional terhadap 19 strain Vibrio parahaemolyticus yang menunjukkan tingkat concordance kategorial sempurna (100%) untuk Oxytetracycline dan Tetracycline, serta 94,12% untuk Enrofloxacin tanpa ditemukan adanya major discrepancy maupun very major discrepancy.

Manual disc diffusion antibiotic sensitivity testing has limitations in terms of technical standardization, accuracy, and operational efficiency. This study aims to design, evaluate the mechanical performance, and validate a semi-automated antibiotic sensitivity testing device based on the EUCAST disk diffusion method to detect Vibrio parahaemolyticus resistance to Tetracycline, Oxytetracycline, and Enrofloxacin. The system integrates automated modules for media preparation, bacterial suspension standardization, inoculation, and antibiotic disc placement. The device's performance was evaluated by comparing it to the manual method based on the accuracy of liquid volume, bacterial density readings, culture homogeneity, bacterial distribution on the agar medium, and the accuracy of antibiotic disc placement. Validation was carried out using 19 Vibrio parahaemolyticus strains against Oxytetracycline, Tetracycline, and Enrofloxacin (171 test points), with the parameters of the size of the inhibition zone and interpretation of antibiotic sensitivity, referring to the standard (manual) system. The test results showed that the peristaltic pump had a volume error of <1 xss=removed>Tetracycline (p-value = 0.509) and Enrofloxacin (p-value = 0.803), and a minimal average difference of -0.85 mm for Oxytetracycline (p-value = 0.015). These numerical results are in line with functional validation of 19 Vibrio parahaemolyticus strains which showed a perfect categorical concordance rate (100%) for Oxytetracycline and Tetracycline, and 94.12% for Enrofloxacin without any major or very major discrepancy.

Kata Kunci : otomatis, uji sensitivitas antibiotik, Vibrio parahaemolyticus

  1. S2-2026-548202-abstract.pdf  
  2. S2-2026-548202-bibliography.pdf  
  3. S2-2026-548202-tableofcontent.pdf  
  4. S2-2026-548202-title.pdf