PENGEMBANGAN DOSIMETER REAL-TIME WIRELESS UNTUK SINAR-X DAN GAMMA BERBASIS GEIGER-MULLER
Imelda Zahra Tungga Dewi, Prof. Drs. Gede Bayu Suparta, MS, Ph.D
2025 | Tesis | S2 Ilmu Fisika
Telah
dilakukan penelitian pengembangan prototipe dosimeter real-time wireless
berbasis sensor GM tipe M4011 yang terintegrasi dengan mikrokontroler Teensy
4.1 dan ESP32. Penelitian dilatarbelakangi oleh tingginya pemanfaatan radiasi
sinar-X dan gamma di bidang medis dan industri serta risiko kesehatan akibat
paparan radiasi berlebihan. Tujuan penelitian ini adalah menghasilkan sistem pemantauan
dosis radiasi secara langsung untuk meningkatkan keselamatan pekerja. Prototipe
dirancang portabel, praktis, dan mampu menampilkan dosis radiasi dalam satuan
CPM (Cacah Per Menit), µSv, dan µSv/h melalui antarmuka berbasis Graphical User
Interface (GUI). Metode penelitian meliputi perancangan perangkat keras dan
perangkat lunak, kalibrasi sensor menggunakan Raysafe X2 (sinar-X) dan
Radiation Alert Ranger (gamma), serta pengujian terhadap variasi jarak dan
sumber radiasi seperti Cs-137, Co-60, dan pesawat sinar-X SF100BY. Hasil
penelitian menunjukkan respons linier yang sangat baik dengan koefisien
korelasi Pearson sebesar 0,997 untuk radiasi gamma dan 0,999 untuk sinar-X,
serta tingkat akurasi sebesar 98,85% (gamma) dan 99,21% (sinar-X). Prototipe juga
berhasil memvalidasi prinsip hukum kuadrat terbalik. Tegangan kerja optimal
tabung GM adalah 380–440 V sesuai karakteristik plateau detektor. Penelitian
ini menunjukkan potensi besar prototipe sebagai dosimeter aktif atau
surveymeter untuk mendukung prinsip ALARA (As Low As Reasonably Achievable)
dalam keselamatan radiasi.
A research study was conducted to develop a real-time
wireless dosimeter prototype based on the GM tube type M4011, integrated with a
Teensy 4.1 microcontroller and ESP32. The study was motivated by the extensive
use of X-ray and gamma radiation in medical and industrial fields and the
associated health risks due to excessive exposure. The objective was to create
a real-time and wireless radiation dose monitoring system to ensure worker
safety. The prototype is designed to be portable, user-friendly, and capable of
displaying radiation dose measurements in CPM (Counts Per Minute), µSv, and
µSv/h through a Graphical User Interface (GUI). The research methodology
involved hardware and software design, sensor calibration using the Raysafe X2
(for X-ray) and Radiation Alert Ranger (for gamma radiation), and testing under
various distances and radiation sources such as Cs-137, Co-60, and the SF100BY
X-ray machine. The results demonstrated a highly linear response with Pearson
correlation coefficients of 0.997 for gamma radiation and 0.999 for X-rays, and
measurement accuracies of 98.85% and 99.21%, respectively. The prototype also
successfully validated the inverse square law principle. The optimal operating
voltage of the GM tube was found to be within the 380–440 V range, consistent
with the plateau characteristics of the detector. These findings indicate the
prototype’s strong potential as an active dosimeter or survey meter supporting
the ALARA (As Low As Reasonably Achievable) principle in radiation safety management.
Kata Kunci : Dosimeter real-time wireless, Geiger-Muller, monitoring radiasi, radiasi gamma, sinar-X, surveymeter