RANCANG BANGUN SISTEM PEMANTAU PARAMETER KUALITAS TIDUR OBJEKTIF BERBASIS PHOTOPLETHYSMOGRAPHY PADA JARI TANGAN
MUHAMMAD RAIHAN NICO SUSILO, Prof. Ir. Sunarno, M.Eng., Ph.D.,IPU. ; Ir. Rony Wijaya, S.T., M.Eng., IPM.
2025 | Skripsi | FISIKA TEKNIK
Tidur adalah
keadaan istirahat alami ditandai penurunan kesadaran, aktivitas otot, dan laju
metabolisme, terbagi menjadi 2 fase : Non-Rapid Eye Movement (NREM) dan Rapid
Eye Movement (REM). Fase NREM
ditandai berkurangnya aktivitas saraf dan detak jantung, sedangkan fase REM
ditandai meningkatnya aktivitas otak seperti saat bangun. Parameter kualitas
tidur objektif dilihat dari efisiensi tidur, serta persentase durasi fase
bangun, NREM, dan REM. Metode standar yang biasa digunakan adalah polysomnography (PSG) dengan alat ukur
utamanya electroencephalograph (EEG)
yang valid dan presisi tinggi dengan mengukur gelombang otak, namun metode tersebut
tidak praktis, mahal, dan butuh tenaga ahli. Sehingga diusulkan metode
alternatif yaitu photoplethysmography
pada jari tangan.
Untuk menyelesaikan masalah tersebut
dilakukan metode penelitian eksperimental dan validasi dengan EEG dengan data
pengujian 10 responden. Sistem pemantau parameter kualitas tidur objektif berbasis
photoplethysmography (PPG) pada jari
tangan yang valid dan presisi berhasil dirancang bangun dengan perangkat keras
berupa sensor pulse heart rate
berbasis LED hijau dan fotodiode mode reflektif, di mana sinyal yang dihasilkan
dikonversi menjadi sinyal digital dan diakuisisi oleh
mikroprosesor, kemudian diproses oleh perangkat lunak di personal computer yang
dilengkapi dengan algoritma ekstraksi fitur, estimasi dengan deteksi puncak,
dan fusi dengan penghalusan temporal untuk menghitung laju pernapasan dan variabilitasnya
sebagai dasar perhitungan parameter kualitas tidur objektif, dengan hasil
analisis : nilai pengulangan/repeatability fase bangun = 2,42%; NREM =
2,09%; REM = 0,99%; efisiensi tidur = 2,42%; nilai reprodusibilitas/reproducibility fase bangun = 2,13%;
NREM = 2,16%; REM = 1,69%; efisiensi tidur = 2,13%; dan nilai bias fase bangun = 3,07?ngan LoA 8,31?n
-2,18%; NREM = -1,25?ngan LoA 5,95?n -8,45%; REM = -1,82?ngan LoA
1,71?n -5,35%; efisiensi tidur = -3,07?ngan LoA 2,18?n -8,33%.
Sleep is a natural resting state
characterized by decreased consciousness, muscle activity, and metabolic rate,
divided into two phases: Non-Rapid Eye Movement (NREM) and Rapid Eye Movement
(REM). The NREM phase is marked by reduced neural activity and heart rate,
while the REM phase shows increased brain activity similar to wakefulness.
Objective sleep quality parameters can be assessed based on sleep efficiency
and the percentage duration of wake, NREM, and REM phases. The standard method
typically used is polysomnography (PSG), with the main device being
electroencephalography (EEG), which is highly precise and valid in measuring
brain waves however, this method is impractical, expensive, and requires a sleep
expert. Therefore, an alternative method using photoplethysmography on the
fingertip is proposed.
To address this issue, an experimental research method was conducted and validated using EEG data from 10 respondents. A valid and precise objective sleep quality parameter monitoring system based on fingertip photoplethysmography was successfully designed, utilizing hardware that consists of a green LED pulse heart rate sensor and a reflective photodiode. The generated signals were converted into digital signals and acquired by a microprocessor, then processed by software on a personal computer equipped with feature extraction algorithms, peak detection estimation, and temporal smoothing fusion to calculate respiratory rate and its variability as a basis for objective sleep quality parameter calculations. The analysis results showed the repeatability value of the wake phase = 2.42%; NREM = 2.09%; REM = 0.99%; sleep efficiency = 2.42%; the standard deviation for reproducibility of the wake phase = 3.86%; NREM = 2.16%; REM = 1.69%; sleep efficiency = 2.13%; and the wake phase bias value is 3.07% with LoA of 8.31% and -2.18%; NREM is -1.25% with LoA of 5.95% and -8.45%; REM is -1.82% with LoA of 1.71% and -5.35%; sleep efficiency is -3.07% with LoA of 2.18% and -8.33%.
Kata Kunci : Tidur, Kualitas Tidur, Photoplethysmography, Jari Tangan