ANALISIS PENGARUH KEMIRINGAN SUDUT TERHADAP HASIL KALIBRASI DIAL INDICATOR MENGGUNAKAN CALIBRATION TESTER BERDASARKAN METODE JAPANESE INDUSTRIAL STANDARD (JIS) B 7503 - 1997

MUHAMMAD MALIK AMIN, Galih Setyawan, S.Si., M.Sc.

2018 | Tugas Akhir | D3 METROLOGI DAN INSTRUMENTASI

Kalibrasi dial indicator dengan variasi kemiringan sudut dial indicator terhadap bidang kerja calibration tester bertujuan untuk mengetahui seberapa besar pengaruh kesalahan geometris (kemiringan sudut) terhadap hasil kalibrasi dial indicator tersebut. Perbandingan hasil kalibrasi dari tiga variasi kemiringan sudut yang telah ditentukan, yaitu sudut 90 derajat, 88 derajat dan 86 derajat berupa error dan hasil ketidakpastian. Metode yang digunakan dalam penelitian ini berdasarkan standar acuan Japanese Industrial Standard (JIS) B 7503 tahun 1997 dengan perbandingan langsung antara dial indicator dan calibration tester. Dial indicator yang digunakan adalah tipe analog merk TOKI yang memiliki kapasitas 10 mm dengan resolusi 0,01 mm. Adapun calibration tester yang digunakan sebagai alat ukur standar adalah merk Mitutoyo tipe 170 102 (analog) dengan kapasitas 25 mm dan resolusi 0,001 mm. Pengukuran dilakukan sebanyak 3x (tiga kali) pengulangan pada pengukuran up dan down pada setiap kemiringan sudut yang telah ditentukan. Hasil kalibrasi dial indicator pada kemiringan sudut 90 derajat diperoleh nilai koreksi rata rata terbesar senilai 0,004 mm untuk pengukuran up dan 0,003 mm untuk pengukuran down. Pada kemiringan sudut 88 derajat diperoleh nilai koreksi rata rata terbesar senilai 0,013 mm untuk pengukuran up dan 0,014 mm untuk pengukuran down. Pada kemiringan sudut 86 derajat diperoleh nilai koreksi rata rata terbesar senilai 0,027 mm untuk pengukuran up dan 0,028 mm untuk pengukuran down. Hasil ketidakpastian bentangan (U95) senilai 5,9 mikrometer pada sudut 90 derajat, sebesar 6,5 mikrometer pada kemiringan sudut 88 derajat dan sebesar 7,7 mikrometer pada kemiringan sudut 86 derajat. Kemiringan sudut berpengaruh signifikan terhadap ketidakpastian kalibrasi dial indicator. Semakin besar kemiringan sudut, maka nilai error dan nilai ketidakpastiannya akan semakin besar juga. Pengaruh variabel kemiringan sudut (x) terhadap variabel ketidakpastian pengukuran (y) sebesar 97,2%. Adapun sisanya sebesar 2,8% dipengaruhi oleh faktor lain yang tidak diikutsertakan dalam penelitian.

Calibration of dial indicator with various angle slope of dial indicator toward calibration tester aims to find out how big the influence of geometric errors (angle of slope) toward calibration result of dial indicator. The comparison result of calibration from three variation of slope angle that has been determined, which is 90 degrees, 88 degrees and 86 degrees in the form of both error and uncertainty result. Method used in this research is based on the reference standard of Japanese Industrial Standard (JIS) B 7503 in 1997 with direct comparison between dial indicator and calibration tester. Dial indicator used is analogue type of brand TOKI with capacity of 10 mm and resolution of 0,01 mm. The calibration tester used as for standard measurement instrument was Mitutoyo brand 170 102 type (analogue) with capacity of 25 mm and resolution of 0,001 mm. Measurement is conducted three times repetition for up and down measurement in each angle of slope that has been determined. Calibration result of dial indicator on slope of 90 degrees obtains the average correction value of 0,004 mm for up measurement and 0,003 mm for down measurement. The slope of 88 degrees obtains the biggest average of correction value of 0,013 mm for up measurement and 0,014 mm for down measurement. The biggest average of correction value on slope 86 degrees is 0,027 mm for up measurement and 0,028 mm for down measurement. The result of uncertainty stretch (U95) is 5,9 micrometre on the slope angle of 90 degrees, 6,5 mikrometre on the slope angle of 88 degrees and 7,7 micrometre on the slope angle of 86 degrees. The angle of slope has significant influence toward uncertainty calibration of dial indicator. The bigger the slope, the bigger the error value and uncertainty value will be. The influence of slope variable (x) toward the variable of uncertainty measurement (y) is 97,2 %. The rest is 2,8 % which is influenced by other factor excluded in the research.

Kata Kunci : dial indicator, calibration of dial indicator, geometric errors, measurement uncertainty.

  1. D3-2018-384570-abstract.pdf  
  2. D3-2018-384570-bibliography.pdf  
  3. D3-2018-384570-tableofcontent.pdf  
  4. D3-2018-384570-title.pdf  
  5. D3-2019-384570-abstract.pdf  
  6. D3-2019-384570-bibliography.pdf  
  7. D3-2019-384570-tableofcontent.pdf  
  8. D3-2019-384570-title.pdf