Laporkan Masalah

Analisa Perbandingan Suhu dan Tegangan Termokopel Probe dan Termokopel Non Probe Tipe K

FIRDA AYU SAKINAH, Galih Setyawan.,M.Sc

2018 | Tugas Akhir | D3 METROLOGI DAN INSTRUMENTASI SV

Telah dilakukan penelitian tentang analisa perbandingan suhu dan tegangan termokopel probe dan termokopel non probe tipe K. Tujuan penelitian ini yaitu untuk mengetahui perbedaan hasil suhu dan tegangan yang dihasilkan dari kedua termokopel tersebut, dan faktor yang mempengaruhi perbedaan hasil termokopel probe dan termokopel non probe. Metode yang digunakan dalam penelitian ini yaitu dengan metode perbandingan langsung suhu standar dan suhu uji dengan penetapan titik-titik suhu menggunakan suhu pemanas Dry Block Calibrator, kemudian alat yang diuji dibandingkan langsung dengan alat standar PRT + PC ETS 10/838769. Berdasarkan penelitian diperoleh selisih nilai suhu terendah yang dihasilkan antara termokopel non probe dan termokopel probe yaitu pada suhu 100oC sebesar 2,3oC. Sedangkan selisih suhu terbesar yaitu 3,69oC pada suhu 250 oC. Hasil uji statistik pada termokopel probe dan termokopel non probe yaitu rata-rata error pada termokopel non probe sebesar 1,76% dengan standar deviasi 0,29 lebih tinggi jika dibandingkan dengan rata-rata error termokopel probe sebesar 1,37% dengan standar deviasi 0,63.Ketidakpastian Bentangan (U95) pada termokopel non probe dan termokopel probe memiliki Bentangan (U95) yang relatif kecil yaitu pada suhu 100 oC, 110 oC,130 oC,140 oC, 180 oC, 190 oC, dan pada suhu 220 oC yaitu 0,00000016. Kata Kunci: Termokopel, Alat Standar PC ETS 10/838769, Dry Block Calibrator.

The research about temperature and valtation comparison in probe thermocouple and non probe thermocouple has been done.The purpose of this research are to know the result difference between the temperature and voltation that is produced by both thermocouples. The method of this research is the straight comparison method between standart temperature and tested temperature with the spot temperature determined by dry block calibrator and the tested device is compared to standart device PRT + PC ETS 10/838769.The result showed that the least temperature difference of probe and non probe thermocouple was 2,3oC at tested temperature of 100 oC,meanwhile the highest temperature was 3,69oC at 250oC temperature. The statistic test results at probe thermocouple and non probe thermocouple was about 1,76% at average with the deviation standart 0,29 which was higher than probe thermocouple error average 1,37% with deviation standart 0,63. The expanded uncertainty(U95) in the non probe thermocouple was relatively smaller at 100 oC, 110 oC,130 oC,140 oC, 180 oC, 190 oC, and 220 oC tempraturewhich was 0,00000016. Keywords: Thermocouple, standart device PC ETS 10/838769, Dry Block Calibrator.

Kata Kunci : Termokopel, Alat Standar PC ETS 10/838769, Dry Block Calibrator


    Tidak tersedia file untuk ditampilkan ke publik.